RFD-200 Series 2
PCI-600
Transformer Turns Ratio Meters
Circuit Breaker Timers
Insulation Resistance Meters
Low Resistance Meters
Current Transformer Testers
Relay Test Equipment
Vacuum Bottle Testers
TYPE
Portable relay test set
INPUT POWER
100 – 120 Vac or 200 – 240 Vac (factory pre-set), 50/60 Hz
PHYSICAL SPECIFICATIONS
16.8”W x 12.6”H x 12”D (42.7 cm x 32.0 cm x 30.5 cm); Weight: 35 lbs (15.9 kg)
AMPERE METER INPUT RANGES
Internal input range: 0 – 250 A; Accuracy: 2% of reading ± 10 mA,
External AC input range: 0 – 6A; Accuracy: 1% of reading ± 10mA
External DC input range: 0 – 6A; Accuracy: 0.5% of reading ±1 count
Measurement Method: True RMS for AC
VOLT METER INPUT RANGE
0 – 600.0V; AC accuracy: 1% of reading ±1 count; DC accuracy 0.5% of reading ±1 count
Measurement Method: True RMS for AC
AUXILIARY POWER SUPPLIES
24 Vdc @ 1 A, 124 Vdc @ 0.25 A, 240 Vac @ 0.125 A
SAFETY
Designed to meet IEC61010 (1995), UL61010A-1, CSA-C22.2 standards
ENVIRONMENT
Operating: -10˚C to 50˚ C (15˚F to +122˚ F); Storage: -30˚ C to 70˚ C (-22˚F to +158˚ F)
OPTIONS
Transportation case
WARRANTY
One year on parts and labor
AC Current Output
Range
No-Load Voltage
Load Voltage
Load Current
Load/Unload Times
10 A
90 Vac
75 Vac
10 A
2 min / 15 min
40 A
25 Vac
20 Vac
40 A
1 min / 15 min
100 A
10 Vac
7.25 Vac
100 A
1 min / 15 min
100 A
10 Vac
3 Vac
250 A
1 sec / 5 min
AC Voltage Output
Range
No-Load Voltage
Load Voltage
Load Current
Load/Unload Times
250 Vac
260 Vac
240 Vac
3 A
10 min / 45 min
DC Voltage Output
Range
No-Load Voltage
Load Voltage
Load Current
Load/Unload Times
300 Vdc
300 Vac
250 Vdc
2 A
10 min / 45 min
Maximum Current
0.4 A
RFD-200 Timer Second Display In seconds and cycles (50/60 Hz programmable)
Range
Resolution
Accuracy
0 to 9.999 sec
1 ms
± (1 ms + 0.01%)
10.00 to 99.99 sec
10 ms
± (10 ms + 0.01%)
100.0 to 999.999 sec
100 ms
± (100 ms + 0.01%)
Timer Cycles Display
Range
Resolution
Accuracy
0 to 9.999 cycles
0.1 cycle
± (0.1 cycle + 0.01%)
1,000 to 49,999 cycles / 50 Hz
1 cycle
± (1 cycle + 0.01%)
The RFD-200 Digital Timer can be started or stopped in three different modes:
Internal Trigger: Start timer with the presence or removal of RFD-200 voltage or current source.
Dry Contact: Change of state of a dry contact input.
Wet Contact: Change of state of a DC voltage input.
Specifications are valid at nominal voltage and ambient temperature of +25˚C (+77˚F).
Last updated on February 16, 2009