Product Search: Results per page: Match: any search words all search words
TYPE Insulation Resistance Meter
PHYSICAL SPECIFICATIONS 19”W x 7H x 15”D, (48 cm x 17 cm x 38 cm)
Weight 24lbs (10.9 kg)
INPUT POWER 110-120 Vac 50/60Hz or 220-240 Vac 50/60Hz, (selectable)
RESISTANCE READING RANGE
(0°C to +30°C)
100 K-ohm to 1 M-ohm ±20%
1 M-ohm to 1 T-ohm ±5%
1 T-ohm to 5 T-ohm ±20% , 5kV
100 K-ohm to 1 M-ohm ±20%
1 M-ohm to 100 G-ohm ±5%
100 G-ohm to 500 G-ohm ±5%, 500 V
1 M-ohm to 10 G-ohm ±5%, 50 V
TEST VOLTAGE Selectable from 50V to 5 kVdc, in 2 Vdc step
OUTPUT VOLTAGE ACCURACY
(0°C to +30°C)
±2% ±1v of selected voltage with load resistance >100 M
SHORT CIRCUIT CURRENT 2mA max
CURRENT READING RANGE 0.03 mA to 2 mA; Accuracy ±5% ±0.2 mA (0°C to +30°C)
CAPACITANCE READING RANGE 0.01 μF to 10.0 μF (Test voltage >100V), Accuracy ±5% ±0.03 μF (0°C to +30°C)
VOLTAGE READING RANGE 50-1250 Vac (rms) or dc, Accuracy ±5%, ±1V (0°C to +30°C)
CAPACITOR DISCHARGE <2 Sec/μF, Automatic at the end of test.
HUM REJECTION 1mA per 1kv of Fast Voltage, 2mA rms maximum
BATTERIES Two 12V, 2.0 Ah Sealed, Lead Acid batteries. Battery life: typical 6 hrs, continuous testing.
The IRM-5000P can be used while charging
SAFETY UL Certified (UL 61010A-1), CAN/CSA Certified (C22.2 No. 1010.1-92)
ENVIRONMENT Operating: -10° to 50° C (15° to +122° F)
Storage: -30° C to 70° C (-22° to +158° F)
CABLES One 6’ cable set, One 15’ cable set, ground cable, power cord and cable bag
OPTIONS Transportation Case
WARRANTY One-Year on Parts and Labor